化学
解吸电喷雾电离
色谱法
代谢物
分析物
质谱法
萃取电喷雾电离
毛细管电泳-质谱法
质谱成像
电喷雾
解吸
代谢组学
电喷雾电离
生物利用度
检出限
原位
溶剂
分析化学(期刊)
串联质谱法
环境电离
作者
Marta Berzaghi,Pedro Ramon Mena,Francesco Potì,Claudio Curti,Daniele Del Rio,Letizia Bresciani
出处
期刊:Food Chemistry
[Elsevier BV]
日期:2026-07-10
卷期号:525 (Pt 1): 150374-150374
标识
DOI:10.1016/j.foodchem.2026.150374
摘要
Desorption electrospray ionization mass spectrometry imaging (DESI-MSI) provides spatial visualization of compounds directly desorbing from tissue surfaces, although detection efficiency is strongly influenced by instrumental settings, tissue characteristics, and analyte properties. This study optimized key DESI XS ion source parameters, including capillary voltage, heat transfer line temperature, and solvent composition to enhance detection of flavan-3-ol metabolites in biologically relevant tissue mimetic models. Eight mimetic models prepared from animal organs were spiked with (poly)phenol metabolite standards at different concentrations and analyzed under multiple parameter combinations. Results highlighted marked tissue- and metabolite-dependent differences in detection efficiency, with some metabolites remaining undetectable. Using optimized conditions hydroxyphenyl-γ-valerolactone-sulfate-d 2 was visualized in liver and kidney tissues following oral gavage of CAT-d 4 in mice. This work emphasizes the importance of tailored DESI-MSI parameter optimization to achieve reliable detection of (poly)phenol metabolites with improved sensitivity, thereby supporting a more accurate in situ characterization of their bioavailability and metabolism.
科研通智能强力驱动
Strongly Powered by AbleSci AI