Alexander Scherl,Yihsuan S. Tsai,Scott A. Shaffer,David R. Goodlett
出处
期刊:Proteomics
[Wiley]
日期:2008-07-01
卷期号:8 (14): 2791-2797
被引量:41
标识
DOI:10.1002/pmic.200800045
摘要
Abstract Although mass spectrometers are capable of providing high mass accuracy data, assignment of true monoisotopic precursor ion mass is complicated during data‐dependent ion selection for LC‐MS/MS analysis of complex mixtures. The complication arises when chromatographic peak widths for a given analyte exceed the time required to acquire a precursor ion mass spectrum. The result is that many measured monoisotopic masses are misassigned due to calculation from a single mass spectrum with poor ion statistics based on only a fraction of the total available ions for a given analyte. Such data in turn produces errors in automated database searches, where precursor m / z value is one search parameter. We propose here a postacquisition approach to correct misassigned monoisotopic m / z values that involves peak detection over the entire elution profile and correction of the precursor ion monoisotopic mass. As a result of using this approach to reprocess shotgun proteomic data we increased peptide sequence assignments by 10% while reducing the estimated false positive ratio from 1 to 0.2%. We also show that 4% of the salvaged identifications may be accounted for by correction of mixed tandem mass spectra resulting from fragmentation of multiple peptides simultaneously, a situation which we refer to as accidental CID.