结构光
投影机
校准
光学
计算机科学
斑点图案
结构光三维扫描仪
计算机视觉
人工智能
数字光处理
像素
物理
量子力学
扫描仪
作者
Canyu Zhu,Qingchuan Zhang
出处
期刊:Applied Optics
[Optica Publishing Group]
日期:2022-08-25
卷期号:61 (27): 8050-8050
被引量:2
摘要
Structured light is a non-contact three-dimensional shape measurement method. The structured light system based on diffractive optical elements (DOEs) is widely used due to its low cost and compact structure. However, compared with a time sequence coded structured light system based on a digital projector, its projector cannot change the content, so it cannot be calibrated with a phase-shift-based method. This paper proposes a calibration method based on the pseudo-camera method and digital image correlation (DIC). It is suitable not only for the calibration of systems based on time series coding but also for the calibration of a speckle structured light system based on DOEs. Validation experiments were conducted in which our method achieved a reprojection error of 0.68 pixels in calibration and a radius error of less than 1% in a measurement of a cylinder in a 20cm×10cm field. Compared with existing calibration methods, this method does not need to set a datum plane, takes fewer photos, operates simply, and has higher calibration efficiency.
科研通智能强力驱动
Strongly Powered by AbleSci AI