Joohee Kim,Jihun Park,Young‐Geun Park,Eunkyung Cha,Minjae Ku,Hyeon Seok An,Kyoung‐Pil Lee,Man‐Il Huh,Junmo Kim,Taek‐Soo Kim,Dai Woo Kim,Hong Kyun Kim,Jang‐Ung Park
标识
DOI:10.1038/s41551-021-00719-8
摘要
Continuous detection of raised intraocular pressure (IOP) could benefit the monitoring of patients with glaucoma. Current contact lenses with embedded sensors for measuring IOP are rigid, bulky, partially block vision or are insufficiently sensitive. Here, we report the design and testing in volunteers of a soft and transparent contact lens for the quantitative monitoring of IOP in real time using a smartphone. The contact lens incorporates a strain sensor, a wireless antenna, capacitors, resistors, stretchable metal interconnects and an integrated circuit for wireless communication. In rabbits, the lens provided measurements that match those of a commercial tonometer. In ten human participants, the lens proved to be safe, and reliably provided accurate quantitative measurements of IOP without inducing inflammation.