James Ronald,Charles Kim,Meghana Konanur,Rui Dai,Nicholas T. Befera,James Knutson,B. Cline
标识
DOI:10.1016/j.jvir.2022.08.030
摘要
A technique to create a coaxial, self-expanding stent graft inside a constraining, bare-metal, balloon-expandable stent for transjugular intrahepatic portosystemic shunt (TIPS) reduction is described. The key steps are performed on a back table rather than inside the patient, and the resulting construct is deployed using standard unsheathing maneuvers. The construct was used in 4 patients to make 6 TIPS diameter reductions (mean postreduction diameter, 6 mm; range, 0-8 mm), all resulting in increases in the portosystemic pressure gradient (mean increase, 6 mm Hg; range, 1-19 mm Hg). On average, hepatic encephalopathy improved 1 point on the West Haven scale (range, 0-2).