植入
医学
股骨颈
聚乙烯
全髋关节置换术
股骨头
外科
火炬
牙科
断裂(地质)
全髋关节置换术
口腔正畸科
材料科学
复合材料
骨质疏松症
内分泌学
物理
天体物理学
作者
Adam P. Breceda,Michael J. DeRogatis,Patrick Lee,Vincent J. Vigorita,E Bryk,Paul S. Issack
标识
DOI:10.1615/jlongtermeffmedimplants.2018025434
摘要
We present the case of a 38-year-old man who presented 7 years after primary total hip replacement with a fracture of the neck of a lateral flare femoral stem and catastrophic polyethylene wear. The unique design of the lateral flare hip stem has been shown previously to be associated with accelerated polyethylene wear, whereas the stem remains well fixed. The resultant polyethylene wear results in the harder cobalt chrome head abrading the titanium shell generating metallic debris. This may have resulted in proximal migration of the implant neck into the cup with subsequent neck-cup impingement and implant fracture.
科研通智能强力驱动
Strongly Powered by AbleSci AI