Jun Li,Wei Chen,Wenli Zhao,Hu Zhang,Zeyong Huang,Shuping Zhang,Yuebing Li
标识
DOI:10.1016/j.cmpb.2019.105238
摘要
Abstract Background The purpose of our study is to compare the reliability and consistency of three commonly used techniques, which are surface measurement, intracardiac electrocardiogram (IECG) and tracheal bifurcation methods in measuring the tip location of totally implantable venous access port (TIVAP) in the same patient. Methods The thirty-five patients scheduled for implantation of TIVAP were included and right subclavian vein was selected for venous access. We used surface measurement method (Group L) to estimate the length and position of catheter before catheterization, and then we used IECG method (Group E) to confirm the position of catheter during catheterization and after catheterization used tracheal bifurcation method by CT (Group T) for measurement. The differences of catheter length measured by three methods were compared by non-parametric Kruskal–Wallis test. Intraclass correlation coefficient (ICC) was used to evaluate the reliability of three methods and Bland-Altman chart was used to evaluate consistency. Results There are no difference in the length of TIVAP catheter between L and T groups (P > 0.05), but they have significant differences comparing with E group (P Conclusions All three methods can be used to measure the length of TIVAP catheter, and IECG method is the most accurate. And similar to most studies, returning the catheter 20 mm after positioning can avoid catheter entering the right atrium, but we found that 10 mm is a sufficient length to achieve this.