高温计
热电偶
发射率
光学
材料科学
分光计
温度测量
校准
微尺度化学
物理
复合材料
数学
量子力学
数学教育
作者
D Keith Coffman,Khalid Hattar,Jian Luo,Shen J. Dillon
摘要
Recent work in ultra-high temperature in situ electron microscopy has presented the need for accurate, contact-free temperature determination at the microscale. Optical measurement based on thermal radiation (pyrometry) is an attractive solution but can be difficult to perform correctly due to effects, such as emissivity and optical transmission, that must be accounted for. Here, we present a practical guide to calibrating and using a spectral pyrometry system, including example code, using a Czerny-Turner spectrometer attached to a transmission electron microscope. Calibration can be accomplished using a thermocouple or commercial heated sample holder, after which arbitrary samples can be reliably measured for temperatures above ∼600∘C. An accuracy of 2% can be expected with the possibility of sub-second temporal resolution and sub-Kelvin temperature resolution. We then demonstrate this capability in conjunction with traditional microscopic techniques, such as diffraction-based strain measurement for thermal expansion coefficient, or live-video sintering evolution.
科研通智能强力驱动
Strongly Powered by AbleSci AI