Jatinder Pal Singh,Neha Batra,Monika Tomar,Arijit Chowdhuri
标识
DOI:10.1149/1945-7111/ae67a8
摘要
Bilirubin (BR) is a key biomarker for haemolytic disorders, with levels >2.5 mg dl −1 indicating hyperbilirubinemia. Accurate and sensitive detection of BR, therefore, assumes importance in clinical diagnostics. This study investigates radio-frequency (RF) sputtered MoO 3 thin films as a matrix for sensitive BR sensors, deposited on indium tin oxide (ITO) coated glass under varying Ar (50%–80%) in the sputtering gas mixture. The film deposited at 65% Ar ambient was noted to exhibit optimal electrochemical characteristics. The same was functionalized with bilirubin oxidase (BOx) via EDC–NHS coupling to develop a MoO 3 -based bioelectrode. Cyclic voltammetry (CV) measurements demonstrated a sensitivity of 50.55 μA (mg/dl) −1 in the linear range of 0.05–2.5 mg dl −1 and a sensitivity of 11.30 μA (mg/dl) −1 in the linear range of 2.5–7.5 mg dl −1 . Differential Pulse Voltammetry (DPV) studies further enhanced detection performance, exhibiting sensitivities of 83.27 μA (mg/dl) −1 and 48.19 μA (mg/dl) −1 in the ranges of 0.05–2.5 mg dl −1 and 2.5–7.5 mg dl −1 , respectively, showing a limit of detection of 0.03 mg dl −1 . The developed BOx/MoO 3 /ITO bioelectrode demonstrated excellent reproducibility, stability and selectivity, confirming the potential of RF-sputtered MoO 3 thin films for reliable bilirubin biosensing applications.