多巴胺
水平扫描速率
电极
电化学
检出限
化学
再现性
循环伏安法
分析化学(期刊)
伏安法
阳极
玻璃碳
盐酸盐
核化学
缓冲溶液
色谱法
有机化学
内科学
医学
物理化学
作者
Deqian Huang,Cheng Chen,Yiming Wu,Hong Zhang,Liang-Quan Sheng,Huajie Xu,Zhao‐Di Liu
标识
DOI:10.1016/s1452-3981(23)19638-6
摘要
In this study, the pretreated glassy carbon electrode (GCE) was prepared by electrochemical oxidation firstly at +1.75 V for 300 s, and then by electrochemical reduction at –1.75 V for 300 s in 0.1 mol L–1 pH 7.0 phosphate buffer solution (PBS) and used for the determination of dopamine (DA) by cyclic voltammetry (CV). The pretreated GCE gives 100-fold greater current responses for dopamine compared with unpretreated GCE. The effect of pH, pretreated mode, scan rate and concentration of dopamine on the peak current was investigated, and the results indicated that the peak current of dopamine is the highest in 0.1 mol L–1 pH 7.0 PBS and the electrode reaction corresponds to a rate-controlled process. The peak current of the anodic peak and the concentration of dopamine hydrochloride is linear in the range of 1.0 × 10–7 – 9.0 × 10–6 mol L–1 and 1.2 × 10–5 – 8.0 × 10–5 mol L–1 with correlation coefficients of 0.9973 and 0.9980, and the detection limit is 3.0 × 10–8 mol/L. It has been successfully applied to the determination of dopamine in dopamine hydrochloride injection with recoveries ranging from 98 to 103%. The proposed method possesses the distinct advantages of simple, appropriate for operation, good reproducibility and cheap instrument.
科研通智能强力驱动
Strongly Powered by AbleSci AI