材料科学
光电探测器
荧光
光学
光电子学
检出限
化学
物理
色谱法
作者
Siying Chen,Xianda Du,He Chen,Pan Guo,Yinchao Zhang,Huiyun Wu
出处
期刊:Applied Optics
[Optica Publishing Group]
日期:2022-05-20
卷期号:61 (17): 5067-5067
被引量:3
摘要
A 64-channel detection system for laser-induced fluorescence (LIF) detection at the cell level is established and applied to single event counting. Generally, fluorescence detection at the cellular level requires a dyeing label to enhance the scattered light intensity for the photodetector. However, the dyeing labels, such as fluorophores, probes, and dyes, complicate sample preparation and increase cytotoxicity in the process. Therefore, label-free detection becomes essential for in vivo research. The presented 64-channel detection system is designed for label-free detection with the ability to record feeble emissions. Two linear photodetector devices are included in the system, extending the wavelength detection range to 366–680 nm with an improved spectral resolution at an average of 4.9 nm. The performance of the system was validated by detecting unlabeled human hepatocytes (L-02) and other cell-level biologic samples. In addition, the 64-channel detection system was also used for particle counting with a quartz microfluidic chip. The counting method is based on fluorescence spectra differs from those of other devices (i.e., flow cytometry and cell-sorting equipment), which use isolated irradiance intensities.
科研通智能强力驱动