Abstract Currently the accepted method for setting the correct rf power levels to achieve 90° and 180° rf pulses for MR imaging is to peak the echo amplitude of a rf spin‐echo sequence. The echo amplitude of this α‐2α pulse sequence is proportional to sin 3 (α) and has a relatively broad maximum. Recently another method for setting the rf flip angle by maximizing the ratio of the stimulated echo to the primary echo amplitudes (in a 3α sequence) demonstrated accuracy similar to that of the spin‐echo method using a shorter repetition time. We present a new, more sensitive, and more accurate method for setting the correct rf power levels for 90° and 180° rf pulses. In this method, based upon the stimulated echo pulse sequence, we are able to accurately set the rf power to within ±0.1 dB by minimizing the signal amplitude of the third spin echo. This null method works for both selective and nonselective rf pulses of flip angle 90° or 180°, allowing the user to accurately adjust the relative amplitudes of the four rf pulse types within a single pulse sequence. © 1989 Academic Press, Inc.