Hannes Deutschmann,M. Augustin,Josef Simbrunner,B Unger,Helmut Schoellnast,Gerald A. Fritz,G.E. Klein
出处
期刊:PubMed
[National Institutes of Health]
日期:2007-04-01
卷期号:28 (4): 628-34
被引量:54
摘要
BACKGROUND AND PURPOSE: 3D time-of-flight MR angiography (3D TOF MRA) may be used as noninvasive alternative to digital subtraction angiography (DSA) for the follow-up of patients with intracranial aneurysms treated with Guglielmi detachable coils (GDCs). We aimed to determine the influence of aneurysm size and location on diagnostic accuracy of 3D TOF MRA for follow-up of intracranial aneurysms treated with GDCs. MATERIALS AND METHODS: Two hundred and one 3D TOF MRAs in 127 consecutive patients with 136 aneurysms were compared with DSA as standard of reference. Sensitivity and specificity of 3D TOF MRA for detection of residual or reperfusion of the aneurysms was calculated with regard to aneurysm size and location. RESULTS: Overall sensitivity and specificity of MRA was 88.5% and 92.9%, respectively. Sensitivity was lower for aneurysms