放气
残余气体分析仪
分压
氢
气体分析仪
材料科学
残余物
分析化学(期刊)
离子
化学
氧气
算法
计算机科学
有机化学
色谱法
环境化学
作者
Shu Watanabe,Masakazu Aono,Shigekí Kato
出处
期刊:Journal of vacuum science & technology
[American Institute of Physics]
日期:1996-11-01
卷期号:14 (6): 3261-3266
被引量:5
摘要
One problem associated with partial pressure measurement in an extreme high vacuum region is outgassing from the residual gas analyzer (RGA) itself. In order to reduce the outgassing rate from the RGA, the temperature of the filament should be lowered, and the heat generated around the ion source of the RGA should be dissipated in the atmosphere. We modified a RGA and reduced its outgassing rate to 1.5×10−12 Pa m3/s. We also determined the partial outgassing rate from the ion source and chamber quantitatively and found that hydrogen constituted more than 99% of the outgassing from the stainless steel.