期刊:Methods in Enzymology
日期:1988-01-01
卷期号:: 351-357
被引量:23
标识
DOI:10.1016/0076-6879(88)58066-7
摘要
Guidelines presented here allow for copper analysis of biological materials by methods that are very sensitive, that require little sample preparation, that have few chemical or spectral interferences, that are inexpensive, and that require only usual care in contamination control. The commercial instruments for FAAS and ETAAS from Perkin-Elmer, from Varian, and from Instrumentation Laboratories Inc. (Allied Analytical Systems) all work well in either the flame or the flameless mode. Background correction techniques are not essential for copper analysis if care is taken with the sample preparation to minimize the background signals. Different types of burners will work adequately if one makes certain that the viscosity of the sample and the control products are similar to the calibration standards. Further, dilution of samples is preferred over increasing the viscosity of the calibration standards by the addition of a protein containing solution or a substance such as glycerol. A 1:10 dilution of blood plasma or serum with dilute nitric acid or water is all that is necessary for copper analysis by the FFAS methods. Cation and anion effects should be tested by bracketing the concentrations of the ions found in the sample with known amounts of ions in the sample solutions. Increasing the concentrations of the ions thought to interfere while keeping the copper concentration constant is another way to test for ion interferences.(ABSTRACT TRUNCATED AT 250 WORDS)
科研通智能强力驱动
Strongly Powered by AbleSci AI