R. Lebert,Christian Pampfer,Andreas Biermanns-Foeth,Thomas Missalla,Christoph Phiesel,Jennifer Arps,Christian Piel
摘要
The scene is set for EUV lithography high volume production. EUV masks with pellicles and scanners with dynamic gas lock thin film windows are considered a corner stone for insertion. For making the thin films usable in EUV scanners, EUV actinic metrology with high precision and accuracy is required to qualify them. One of the key features is the uniformity of the transmission at 13.5 nm through the thin film membrane to be qualified over the entire area. The specification for the required tool is qualifying a full sized EUV pellicle with an average EUV transmission of around 90 % to a uniformity of < 99.6 % – hence precision and accuracy of the process of below 0.1 % are demanded. Certainly, no particles should be added to the EUV pellicle by the qualification measurement. Within RI Research Instrument's support of EUVL infrastructure by developing actinic metrology solutions, we have built a dedicated tool for EUV pellicle transmission characterization using the effective inband EUV scheme. The actinic measurement of the full pellicle is accomplished in less than 40 minutes. Vacuum loading and unloading consumes another 35 minutes due to the need of extreme care for such fragile samples and to avoid contamination. First images of full sized pellicles show precision and accuracy of better 0.1 % for a qualification pixel resolution of 300x300 μm2.