作者
Eraldo Occhetta,Miriam Bortnik,R. Audoglio,C VASSANELLI
出处
期刊:Europace
[Oxford University Press]
日期:2004-11-01
卷期号:6 (6): 538-547
被引量:111
标识
DOI:10.1016/j.eupc.2004.08.009
摘要
To determine whether dual-chamber rate-adaptive Closed Loop Stimulation (CLS) could prevent recurrence of Vasovagal Syncope (VVS). During VVS, an increase in myocardial contractility associated with a reduction of ventricular filling produces an increase in baroreceptor afferent flow and a consequent decrease in the heart rate. The CLS algorithm is a form of rate-adaptive pacing, which responds to myocardial contraction dynamics, by measuring variations in right ventricular intracardiac impedance: during an incipient VVS it could increase paced heart rate and avoid bradycardia, arterial hypotension and syncope. Fifty patients (27 males, mean age 59 ± 18 year) with severe and recurrent vasovagal syncope and positive Head Up Tilt Test (HUTT) with cardioinhibition, received a CLS pacemaker (INOS 2 , Biotronik GmbH Co., Germany). The primary end point was recurrence of two VVSs during a minimum of 1 year of follow-up. Randomization between DDD-CLS and DDI mode (40 bpm) pacing was performed only during the first stage of the study (first year): 9/26 randomized to DDI mode (control group) and 17/26 in DDD-CLS mode. All the 24 patients recruited in the second stage of the study (second year) were programmed in DDD-CLS mode. Of the nine patients randomized to the DDI mode, seven had recurrences of syncope during the first year. At the end of the first year the nine patients were reprogrammed to the CLS mode and no syncope occurred after reprogramming. The 41 patients programmed to CLS had a mean follow-up of 19 ± 4 months: none reported VVS, only four (10%) reported occasional presyncope and their quality of life greatly improved. Positive HUTT at the end of the first year failed to predict the clinical response to CLS pacing. The study demonstrates the effectiveness of CLS pacing in preventing cardioinhibitory VVS. A possible placebo effect of pacemaker implantation occurred in 22% of patients.
科研通智能强力驱动
Strongly Powered by AbleSci AI