Hydroxyapatite columns for high-performance liquid chromatography that are reusable for a long time were developed; performance tests were carried out by using several types of protein. Using a high flow rate, a sharp chromatographic peak can be obtained for a homogeneous molecule. A very high level of chromatographic separation can be achieved by decreasing the slope of the gradient and increasing the total column length at the same time.