Kaicheng Liang,Osman O. Ahsen,Zhao Wang,Hsiang-Chieh Lee,Wenxuan Liang,Benjamin M. Potsaid,Tsung Han Tsai,Michael G. Giacomelli,Vijaysekhar Jayaraman,Hiroshi Mashimo,Xingde Li,James G. Fujimoto,Kaicheng Liang,Osman O. Ahsen,Zhao Wang,Hsiang-Chieh Lee,Wenxuan Liang,Benjamin M. Potsaid,Tsung Han Tsai,Michael G. Giacomelli
出处
期刊:Optics Letters
[The Optical Society]
日期:2017-08-08
卷期号:42 (16): 3193-3193
被引量:39
摘要
Endoscopic optical coherence tomography (OCT) instruments are mostly side viewing and rely on at least one proximal scan, thus limiting accuracy of volumetric imaging and en face visualization. Previous forward-viewing OCT devices had limited axial scan speeds. We report a forward-viewing fiber scanning 3D-OCT probe with 900 μm field of view and 5 μm transverse resolution, imaging at 1 MHz axial scan rate in the human gastrointestinal tract. The probe is 3.3 mm diameter and 20 mm rigid length, thus enabling passage through the endoscopic channel. The scanner has 1.8 kHz resonant frequency, and each volumetric acquisition takes 0.17 s with 2 volumes/s display. 3D-OCT and angiography imaging of the colon was performed during surveillance colonoscopy.