摘要
Nicotine (NCT) is commonly contained in e-cigarette oils, and long-term nicotine intake can exacerbate cardiovascular disease, neur odegenerative disease, and lung cancer. Therefore, there is a need to develop a fast and sensitive method for detecting NCT in e-cigarette oils. We prepared anti-NCT monoclonal antibodies (mAbs), which belong to the IgG2 subclass. The optimum mAb had a half-maximal inhibitory concentration (IC 50 ) of 6.81 ng/mL for NCT. We developed a colloidal gold immunochromatographic strip (CG-ICS) to detect NCT content with a visual cut-off and limit of detection of 20 ng/mL and 0.40 ng/mL, respectively. The recovery rates of NCT from spiked samples ranged between 100.36% and 109.96% based on CG-ICS. These results were consistent with those obtained from high performance liquid chromatography (HPLC). We confirmed the reliability of CG-ICS using NCT-positive samples. Therefore, CG-ICS represents an effective and rapid on-site screening tool for the detection of NCT in e-cigarette oils. • Anti-NCT mAbs were prepared, which had a IC 50 of 6.81 ng/mL for NCT. • CG-ICS was developed to detect NCT content with a visual cut-off and LOD of 20 ng/mL and 0.40 ng/mL, respectively. • The recovery rates of NCT from spiked samples ranged between 97.42% and 103.93% based on CG-ICS. • These results were consistent with those obtained from HPLC.