医学
结肠镜检查
质量评定
医学物理学
前瞻性队列研究
质量(理念)
物理疗法
梅德林
定量评估
风险评估
重症监护医学
生活质量(医疗保健)
质量管理
试验预测值
疾病严重程度
患者评估
临床实习
心理测量学
质量保证
作者
Todd A. Brenner,Chris Labaki,Joseph D. Feuerstein,Tyler M. Berzin
标识
DOI:10.14309/ajg.0000000000003855
摘要
INTRODUCTION: We report a prospective pilot of GI Genius ColonPro 4.0 (Medtronic, Minneapolis, MN), the first US Food and Drug Administration-approved computer-aided quality (CAQ) platform for colonoscopy. METHODS: ColonPro measurements of withdrawal time, cecal intubation, and Boston Bowel Preparation Score (BBPS) were compared with independent video review in 110 outpatient colonoscopies. RESULTS: Agreement between the CAQ and physicians was poor for BBPS (intraclass correlation coefficient = 0.40, P < 0.001) but strong for withdrawal time (intraclass correlation coefficient = 0.89, P < 0.001) and cecal intubation (κ = 0.79, P < 0.001). DISCUSSION: CAQ reliably measures withdrawal time and cecal intubation, but artificial intelligence-based BBPS scoring requires further validation before integration into current paradigms of colonoscopy quality.
科研通智能强力驱动