医学
支架
PLGA公司
尿道
膀胱镜检查
增生
生物相容性
外科
聚己二酮
泌尿科
生物医学工程
泌尿系统
解剖
病理
材料科学
纳米颗粒
冶金
纳米技术
作者
Andres Kotsar,Taina Isotalo,Joonas Mikkonen,Hanne Juuti,Paula Martikainen,Martti Talja,Minna Kellomäki,Pertti Törmälä,Teuvo L.J. Tammela
标识
DOI:10.1089/end.2007.0047
摘要
Purpose: The biodegradable PLGA (a copolymer of L-lactide and glycolide) urethral stent with a spiral configuration has been used clinically for the prevention of postoperative urinary retention after different types of thermal therapy for benign prostatic hyperplasia. A new braiding pattern for this stent has recently been developed by our group. The aim here was to investigate the in situ degradation and biocompatibility of the new braided stent in the rabbit urethra. Materials and Methods: PLGA stents with a one-over-one braiding pattern and steel stents served as controls that were inserted into the posterior urethras of 24 male rabbits using a special delivery instrument. The animals were sacrificed after 1 week, 1 month, 2 months, or 4 months, and light microscopy and histologic analyses were performed. Results: The delivery instrument worked well and cystoscopy was not needed in the insertion process. The braided PLGA stents degraded smoothly in 1 to 2 months. The metallic stents induced more epithelial hyperplasia and epithelial changes than the biodegradable stents at all time points analyzed. These differences increased during follow-up. Conclusion: The degradation process was well controlled and the biodegradable stents were more biocompatible than the metallic stents. The new stent can be inserted into the posterior urethra without cystoscopic aid.
科研通智能强力驱动
Strongly Powered by AbleSci AI