Core temperature monitoring is important for the assessment and prevention of possible postoperative complications. The aim of the present study was to examine the agreement between the core temperature values and the forehead, tympanic membrane and axillary values in postoperative adult patients in clinical practice. The study measured the core temperature of 65 patients undergoing scheduled abdominal surgery using SpotOn™ and compared these with those obtained using non-contact forehead infrared, infrared tympanic and axillary thermometers. Correlation and Bland-Altman analyses were conducted for these comparisons. All temperatures were recorded at 4h intervals after postoperative arrival to the ward. Forehead temperature recordings showed a good correlation with the core temperature with excellent accuracy and was comparable to the tympanic temperature. Both forehead and tympanic thermometers can rapidly and effectively measure the core temperature during early postoperative period. Considering patients' safety, non-contact forehead infrared thermometers may be useful for postoperative monitoring.