离子迁移光谱法
葡萄酒
色谱法
重复性
化学
检出限
质谱法
气相色谱-质谱法
气相色谱法
芳香
火焰离子化检测器
分析化学(期刊)
食品科学
作者
Wenyao Zhu,Frank Benkwitz,Bahareh Sarmadi,Paul A. Kilmartin
标识
DOI:10.1021/acs.jafc.1c06411
摘要
A new quantitative method based on static headspace-gas chromatography-ion mobility spectrometry (SHS-GC-IMS) is proposed, which enables the simultaneous quantitation of multiple aroma compounds in wine. The method was first evaluated for its stability and the necessity of using internal standards as a quality control measure. The two major hurdles in applying GC-IMS in quantitation studies, namely, nonlinearity and multiple ion species, were also investigated using the Boltzmann function and generalized additive model (GAM) as potential solutions. Metrics characterizing the model performance, including root mean squared error, bias, limit of detection, limit of quantitation, repeatability, reproducibility, and recovery, were investigated. Both nonlinear fitting methods, Boltzmann function and GAM, were able to return desirable analytical outcomes with an acceptable range of error. Potential pitfalls that would cause inaccurate quantitation, that is, effects of ethanol content and competitive ionization, were also discussed. The performance of the SHS-GC-IMS method was subsequently compared against that of a currently established method, namely, GC-MS, using commercial wine samples. These findings provide an initial validation of a GC-IMS-based quantitation method, as well as a starting point for further enhancing the analytical scope of GC-IMS.