作者
Laura Basiricò,Matteo Verdi,Andrea Ciavatti,Lorenzo Piergallini,Elisa Grassi,Federica Fioroni,Lorenzo Margotti,N. Tosi,Valentina Cicero,A. Montanari,Mauro Iori,Beatrice Fraboni
标识
DOI:10.1002/admt.202401111
摘要
Abstract On‐line in situ monitoring of radiation emitted by radiotracers during intravenous injection is of crucial importance for the safety of patients and personnel involved in nuclear medicine treatments and analysis. The aim of this work is to assess wearable perovskite ionizing radiation detectors for such application. With this purpose pixelated direct detector based on thin films of 2D layered perovskite PEA 2 PbBr 4 are tested for the monitoring of gamma‐rays emitted by 18 F radiopharmaceutical under different experimental conditions. The wearable detector is here demonstrated being able to detect in real‐time the gamma‐rays emitted by annihilation of 18 F radiotracer. Its reliability is assessed by 18 F half‐life measurement with a maximum accuracy deviation within 8%. The detector also successes in its validation for the foreseen application, demonstrating able to monitor in real‐time the whole injection process and extravasation event.