Yoshiaki Morishima,Koichi Chida,Takayoshi Meguro,Morihisa Hirota,Hiroo Chiba,Hiroshi Fukuda
摘要
Abstract This study aimed to compare the lens equivalent dose (LED) measured during endoscopic retrograde cholangiopancreatography (ERCP) using DOSIRIS™ as a dedicated dosemeter to that measured using glass badges to determine if glass badges can be alternative tools for LED measurement. LEDs for physicians during ERCP were measured using the DOSIRIS™ [3-mm dose equivalent] worn on the outer edge of the eyes and personal dosemeters (glass badges) [0.07-mm dose equivalent] worn on the right and left sides of the neck. The cumulated doses over 6 months for the left eye using DOSIRIS™ were 9.5 and 11.8 mSv for physicians A and B, whereas doses measured using glass badges were 7.5 and 11.6 mSv, respectively. The LEDs of the physicians at the left eye and left neck side showed almost similar values and were significantly correlated (r = 0.95; p < 0.01). For an accurate LED measurement during ERCP, using a dosemeter such as DOSIRIS™ is recommended, although similar LED estimation values were reported using glass badges on the left neck side.