Yabin Chen,Jian Huang,Huidan Chen,Ling-Hui Ma,Min Dou,Qiurong Zhao,Zhishan Zhang,Xiaomei Wen
出处
期刊:PubMed
[National Institutes of Health]
日期:2024-09-01
卷期号:54 (5): 671-678
摘要
OBJECTIVE: The purpose of this study is to evaluate the accuracy of Sysmex XN-9000 hematology analyzer (XN-9000) in detecting nucleated cell classifications in serous effusion and to further evaluate whether it can be used for screening malignant serous effusion. METHODS: The consistency between XN-9000 and manual microscopy in the classification of nucleated cells was evaluated using Passing-Bablok regression and Bland-Altman plot. ROC analysis was applied to evaluate the value of HF-BF% in screening malignant specimens. RESULTS: /μL, there was high consistency in detecting NE-BF%, LY-BF%, and HF-BF%. ROC analysis showed that HF-BF% can be used for screening malignant serous effusion, with an area under the curve (AUC) of 0.572 (95%CI=0.504~0.639). CONCLUSION: /μL. The optimal cut-off value is 13.15%.