扫描仪
偏移量(计算机科学)
成像体模
梯度回波
翻转角度
补偿(心理学)
物理
核磁共振
核医学
生物医学工程
材料科学
光学
计算机科学
磁共振成像
放射科
医学
心理学
精神分析
程序设计语言
作者
Xia Zhao,Hyunyeol Lee,Hee Kwon Song,Cheng‐Chieh Cheng,Félix W. Wehrli
摘要
Purpose The impact of gradient imperfections on UTE images and UTE image‐derived bone water quantification was investigated at 3 T field strength. Methods The effects of simple gradient time delays and eddy currents on UTE images, as well as the effects of gradient error corrections, were studied with simulation and phantom experiments. The k‐space trajectory was mapped with a 2D sequence with phase encoding on both spatial axes by measuring the phase of the signal in small time increments during ramp‐up of the read gradient. In vivo 3D UTE images were reconstructed with and without gradient error compensation to determine the bias in bone water quantification. Finally, imaging was performed on 2 equally configured Siemens TIM Trio systems (Siemens Medical Solutions, Erlangen, Germany) to investigate the impact of such gradient imperfections on inter‐scanner measurement bias. Results Compared to values derived from UTE images with full gradient error compensation, total bone water was found to deviate substantially with no (up to 17%) or partial (delay‐only) compensation (up to 10.8%). Bound water, obtained with inversion recovery‐prepared UTE, was somewhat less susceptible to gradient errors (up to 2.2% for both correction strategies). Inter‐scanner comparison indicated a statistically significant bias between measurements from the 2 MR systems for both total and bound water, which either vanished or was substantially reduced following gradient error correction. Conclusion Gradient imperfections impose spatially dependent artifacts on UTE images, which compromise not only bone water quantification accuracy but also inter‐scanner measurement agreement if left uncompensated.
科研通智能强力驱动
Strongly Powered by AbleSci AI