活检
前列腺活检
医学
前列腺癌
临床试验
放射科
机械人手术
外科
癌症
内科学
作者
Wenhe Jiang,Yongzhuo Gao,Mingwei Wen,Qiangqiang Huang,Hui Dong,Huageng Liang,Dongmei Wu,Wei Dong
摘要
Robotic systems are increasingly utilised to improve clinical outcomes in prostate interventions. We enroled 20 patients to verify the clinical feasibility of transrectal ultrasound (TRUS)-guided robot-assisted prostate biopsy (PBx). For each patient, controlled trials of systematic 12-core PBx were initially conducted manually using a biopsy gun, followed by a 3-core cognitive fusion targeted biopsy performed by the hand-held robot. The targeting errors for robot-assisted biopsy and freehand biopsy were 3.71 ± 1.41 mm and 3.94 ± 1.49 mm, respectively. Their cancer core rates were 19.1% and 12.6%, and the average duration of each biopsy was 6.94 min and 1.62 min, respectively. The robot's sampling success rate was lower than that of freehand with a biopsy gun in the clinical trial. However, robot-assisted PBx has shown improved cancer core rate, slightly enhanced surgical accuracy, and the capability to reach clinically significant tumours' 5 mm error range, all demonstrating its clinical value.
科研通智能强力驱动
Strongly Powered by AbleSci AI