透视
医学
介入放射学
股动脉
忠诚
外科
放射科
计算机科学
电信
作者
Péter Legeza,Kalyna Sconzert,John‐Michael Sungur,Thomas M. Loh,Gavin W. Britz,Alan B. Lumsden
摘要
Abstract Background Robotic‐assisted endovascular surgery enables us to perform interventions from long distances. This study evaluates the workflow and telecommunication requirements of telerobotic peripheral vascular interventions. Methods Ten superficial femoral artery cases were performed by the operator being 44 miles away from the interventional suite, with an endovascular robotic system, on a high‐fidelity endovascular simulator. Procedural success, technical success, fluoroscopy time, residual stenosis, contrast dose and network delay were registered. Communication success was assessed after each procedure on a scale from 1 (unacceptable) to 5 (ideal). Results Procedural success and technical success were 100% and 80%, respectively. The mean residual stenosis, fluoroscopy time and contrast dose were 1.7 ± 5.25%, 6.5 ± 1.8 min and 58.8 ± 14.8 ml. The mean network latency was 38.9 ± 3.5 ms. Median communication success scores were 4.5 (min: 4, max: 5) reported by both the operator and the bedside technician on a scale of 1 (unacceptable) to 5 (ideal). Conclusion With a stable network connection and good communication protocol, a high success rate was achieved for remote robotic‐assisted peripheral vascular intervention in an ex vivo model.