Liquid infused surfaces (LIS) hold remarkable potential in anticoagulation. However, liquid loss of LIS in the bloodstream remains a challenge toward its clinical application. Here, micronano hierarchy structures are obtained on the titanium alloy substrate by regulating the microspheres' distribution. When the gap between the microspheres is smaller than the diameter of the red blood cell (RBC), the LIS is more stable under the blood wash and presents a better anticoagulation performance. The proper interval is found to prevent the RBCs from entering the gap and remove the liquid on the surface. The retained thickness of the liquid film is measured by the atomic force microscopy (AFM) technique. The LIS is applied on the front guide vane of an artificial heart pump and exhibits significant improvement on anticoagulation in the blood circulation in vitro for 25 h. The techniques and findings can be used to optimize the anticoagulation performance of LIS-related biomedical implant devices.