医学
全膝关节置换术
关节置换术
外科
骨科手术
膝关节手术
物理医学与康复
骨关节炎
病理
替代医学
作者
Johan Bellemans,Hilde Vandenneucker,Johan Vanlauwe
标识
DOI:10.1097/blo.0b013e318126c0c0
摘要
UNLABELLED: Increasing evidence suggests performing total knee arthroplasty using computer navigation can lead to more accurate surgical positioning of the components and knee alignment compared to a conventional operating technique without computer assistance. The use of robotic technology could theoretically take this accuracy one level further because it uses navigation in combination with ultimate mechanical precision, which could eliminate or reduce the inevitable margin of error during mechanical preparation of the bony cuts of total knee arthroplasty by the surgeon. We prospectively followed 25 consecutive cases using an active surgical robot. The minimum followup was 5.1 years (mean, 5.5 years; range, 5.1-5.8 years). Our results demonstrate excellent implant positioning and alignment was achieved within the 1 degree error of neutral alignment in all three planes in all cases. Despite this technical precision, the excessive operating time required for the robotic implantation, the technical complexity of the system, and the extremely high operational costs have led us to abandon this procedure and direct our interest more toward smart semiactive robotic systems. LEVEL OF EVIDENCE: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.
科研通智能强力驱动
Strongly Powered by AbleSci AI