生物传感器
结核分枝杆菌
检出限
痰
肺结核
纳米技术
材料科学
线性范围
化学
色谱法
医学
病理
作者
Jinbiao Ma,Manman Du,Can Wang,Xinwu Xie,Hao Wang,Tie Li,Shixing Chen,Lixia Zhang,Shun Mao,Xiaohong Zhou,Min Wu
出处
期刊:ACS Sensors
[American Chemical Society]
日期:2021-09-02
卷期号:6 (9): 3367-3376
被引量:43
标识
DOI:10.1021/acssensors.1c01227
摘要
Tuberculosis (TB) mostly spreads from person to person through Mycobacterium tuberculosis (MTB). However, the majority of conventional detection methods for MTB cannot satisfy the requirements for actual TB detection. As one of the most promising powerful platforms, a silicon nanowire field-effect transistor (SiNW-FET) biosensor shows good prospect in TB detection. In this study, an enhanced SiNW-FET biosensor was developed for the rapid and sensitive detection of MTB. The surface functional parameters of the biosensor were explored and optimized. The SiNW-FET biosensor has good sensitivity with a detection limit of 0.01 fg/mL toward protein. The current change value shows a linear upward trend with the increase in protein concentration in the range of 1 fg/mL to 100 μg/mL. One whole test cycle can be accomplished within only 30 s. More importantly, a good distinction was realized in the sputum without pretreatment between normal people and TB patients, which greatly shortened the TB detection time (only 2–5 min, considering the dilution of sputum). Compared with other methods, the SiNW-FET biosensor can detect MTB with a remarkably broad dynamic linear range in a shorter time.
科研通智能强力驱动
Strongly Powered by AbleSci AI