化学
色谱法
硝苯地平
选择性反应监测
甲酸
质谱法
萃取(化学)
检出限
校准曲线
串联质谱法
三级四极质谱仪
电喷雾电离
分析化学(期刊)
钙
有机化学
作者
Pan Xi-gui,Sanming Zhou,Qin Fu,Xianming Hu,Jianhong Wu
摘要
ABSTRACT Nifedipine is a dihydropyridine calcium channel blocker used widely in the management of hypertension and other cardiovascular disorders. In this work, a simple, rapid and sensitive liquid chromatography/tandem mass spectrometry method was developed and validated to determine nifedipine in dog plasma using nimodipine as the internal standard. Chromatographic separation was carried out on a C 8 column. The mobile phase consisted of a mixture of acetonitrile, water and formic acid (60:40:0.2, v/v/v) at a flow rate of 0.5 mL/min. Detection was performed on a triple quadrupole tandem mass spectrometer in selected reaction monitoring mode via an atmospheric pressure chemical ionization source. The method has a lower limit of quantification of 0.20 ng/mL with consumption of plasma as low as 0.05 mL. The linear calibration curves were obtained in the concentration range of 0.20–50.0 ng/mL ( r = 0.9948). The recoveries of the liquid extraction method were 74.5–84.1%. Intra‐day and inter‐day precisions were 4.1–8.8 and 6.7–7.4%, respectively. The quantification was not interfered with by other plasma components and the method was applied to determine nifedipine in plasma after a single oral administration of two controlled‐release nifedipine tablets to beagle dogs. Copyright © 2013 John Wiley & Sons, Ltd.